ISO 21501-3:2019

Ausgabedatum: 2019 11 22

Determination of particle size distribution — Single particle light interaction methods — Part 3: Light extinction liquid-borne particle counter

This document describes a calibration and verification method for a light extinction liquid-borne particle counter (LELPC), which is used to measure the size and particle...
Weiterlesen
Gültig
Herausgeber:
International Organization for Standardization
Format:
Digital | 18 Seiten
Sprache:
Englisch

This document describes a calibration and verification method for a light extinction liquid-borne particle counter (LELPC), which is used to measure the size and particle number concentration of particles suspended in liquid. The light extinction method described in this document is based on single particle measurements. The typical size range of particles measured by this method is between 1 µm and 100 µm in particle size.

The method is applicable to instruments used for the evaluation of the cleanliness of pharmaceutical products (e.g. injections, water for injections, infusions), as well as the measurement of number and size distribution of particles in various liquids.

The following are within the scope of this document:

— size setting error;

— counting efficiency;

— size resolution;

— maximum particle number concentration;

— sampling flow rate error;

— sampling time error;

— sampling volume error;

— calibration interval;

— reporting results from test and calibration.

ISO 21501-3:2019
2019 11 22
Determination of particle size distribution — Single particle light interaction methods — Part 3: Li...
Norm
ISO 21501-3:2007
2007 05 07
Determination of particle size distribution — Single particle light interaction methods — Part 3: Li...
Norm
ISO 13323-1:2000
2000 11 02
Determination of particle size distribution — Single-particle light interaction methods — Part 1: Li...
Norm